ILE 256-PL/H
- Industrial housing IP65
- Air purge for lens
- Water-cooling
- Operation with or without PC
- Fiber optic data transmission
- No opto-mechanical scanner
- Application specific hardware
and software
- 256 data points/line at 256 lines/second
- Uncooled infrared linear array
- 16 bit A/D conversion
- Large dynamic range
- Triggered measurement
- Alarm function
- Data recording
Infrared line camera with 256 data points per line and a measurement rate of 512 lines per second - an industrial infrared camera for non-contact temperature measurement of production processes at temperatures of 100 °C to 800 °C. Monitoring of limits and control of processes is possible with the powerful online software running with Windows®. With the data recording the measured data can be saved as a video.
Spectral range 8 µm to 14 µm
Measurement range1 100 °C to 800 °C Sensor uncooled pyroelectric linear array (128 pixels)
| Lens1 Type |
standard |
option |
| Field of view (FOV) |
40° x 0.3° |
56° x 0.5° |
| Measurement distance |
> 20 cm |
> 50 cm |
Measurement uncertainty2 1 K + 1 % of the measured value in °C
Noise equivalent temperature difference2 < 0.5 K (32 Hz), < 1.5 K (512 Hz)
Frame rate internal 512 Hz, selectable: 512 Hz, 256 Hz, 128 Hz, 64 Hz, 32 Hz, ...
Response time internal 4 ms, selectable: 2 / measurement frequency
Interface3 RS232 (32 Hz max) RS422 (64 Hz max) RS232 fiber optic (32 Hz max) PCI fiber optic (128 Hz max)
Digital inputs (trigger) 2 electrically isolated LED inputs (optocouplers)
Digital outputs (alarm) 2 electrically isolated OC outputs (optocouplers)
Connectors3 round connector with thread interlocking (16 pins) interlocking fiber optic connector (2 fibers) water supply tubing (nominal width 4 mm, 2 bar max) compressed air tubing (nominal width 6 mm, 2 bar max)
Weight appr. 3.2 kg
Power supply 18 V to 36 V DC, 10 VA
Housing protection degree IP65, optional with integrated water-cooling system, air purge, fixed or swivel mounting base
Camera operating temperature 0 °C to 50 °C (without water-cooling) -25 °C to 150 °C (with water-cooling)
System cable operating temperature -25 °C to 150 °C
Fiber optic operating temperature -10 °C to 70 °C
Storage conditions -20 °C to 70 °C, rel. humidity: 95 % max
Software PC control and display program for Windows®
1 Other on request 2 Specification for blackbody reference, ambient temperature 25 °C 3 Depending on configuration
|